HL7 - Version 2.9


Table 0365: Equipment State

Interpretation: Gerätestatus
OID Table: 2.16.840.1.113883.12.365
Codesystem OID: 2.16.840.1.113883.18.223 Version 3
Value Set OID: 2.16.840.1.113883.21.239
Binding: universal
Expansion: all codes from codesystem
Vocabulary Domain: tbd
Table Type: HL7
Case insensitive: Falsch
Steward: OO
Section 2.C.2.290

Value Description German Interpretation Comment Chapter
IN Software is ready, hardware not yet

2.C.2.290
CO Configuring

2.C.2.290
PU Software and hardware are not yet ready

2.C.2.290
RS Software and hardware are ready, but user action is required to start


ID Successfully started, new orders can be accepted, currently no orders are present

2.C.2.290
OP Successfully started, new orders can be accepted

2.C.2.290
CL Clearing

2.C.2.290
PA Pausing

2.C.2.290
PD User action is required to continue

2.C.2.290
ES Error, remaining orders can be finished, new orders cannot be accepted

2.C.2.290
TS Transport stopped

2.C.2.290
SS Sampling stopped

2.C.2.290
SD Shutting down

2.C.2.290
DI Diagnose

2.C.2.290
MA Maintenance

2.C.2.290
FL Failure, remaining orders are aborted, new orders cannot be accepted


UNK Unknown


LT For diagnostic instruments: some test types are unavailable


Comments made by the German HL7 User Group:

Gerätestatus

Used Within the following Data Elements

01323 Equipment State / Gerätestatus (13.4.1.3)